Z-1832-2013 Class II Terminated

Recalled by Exactech, Inc. — Gainesville, FL

FDA device recall Z-1832-2013 was initiated by Exactech, Inc. on May 30, 2013 and is designated Class II. Reason for recall: When torque is applied to the Articulated Driver at the maximum angle, the cross pin may shift out of position, potentially leading to disassembly of the instrument. The recall status is terminated (terminated Sep 19, 2013). Affected quantity: 57.

Recall Details

Product Type
Devices
Report Date
Aug 7, 2013
Initiation Date
May 30, 2013
Termination Date
Sep 19, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
57

Product Description

Exactech Articulated Driver A nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery.

Reason for Recall

When torque is applied to the Articulated Driver at the maximum angle, the cross pin may shift out of position, potentially leading to disassembly of the instrument.

Distribution Pattern

US Distribution including the states of AR, CA, FL, NJ, NY, OK, and TX

Code Information

Lot #: 52667001 Catalog #: 181-00-04

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1832-2013
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.