Z-1838-2024 Class II Ongoing

Recalled by Philips North America Llc — Cambridge, MA

Recall Details

Product Type
Devices
Report Date
May 29, 2024
Initiation Date
May 3, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

Achieva 1.5T Initial system Model Number (REF): 781178

Reason for Recall

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Code Information

UDI-DI: N/A Serial Numbers: 20282, 20303