Z-1844-2024 Class II Ongoing
FDA device recall Z-1844-2024 was initiated by Philips North America Llc on May 3, 2024 and is designated Class II. Reason for recall: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or… The recall status is ongoing.
Recall Details
- Product Type
- Devices
- Report Date
- May 29, 2024
- Initiation Date
- May 3, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
Product Description
Evolution upgrade 3.0T Model Number (REF): (1)782117; (2) 782143
Reason for Recall
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Code Information
UDI-DI: N/A Serial Numbers: 42246 85713 42029 71774 85668 85680 85685 85694 85705 42023 42024 42375 42217 42402 71044 85698 85700 85146 85677 85699 71595 71815 85649 42320 85684 71836 85679 85682 85697 85701 85708 85714