Z-1855-2024 Class II Ongoing

Recalled by Philips North America Llc — Cambridge, MA

FDA device recall Z-1855-2024 was initiated by Philips North America Llc on May 3, 2024 and is designated Class II. Reason for recall: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or… The recall status is ongoing. Affected quantity: 118 units.

Recall Details

Product Type
Devices
Report Date
May 29, 2024
Initiation Date
May 3, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
118 units

Product Description

Panorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781350

Reason for Recall

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Code Information

UDI-DI: (1) 782152 N/A (2) 782110 (01)00884838099364(21) Serial Numbers: Serial Number 37019 37067 37149 37188 37184 37034 37036 37058 37167 37173 37183 37195 37197 37009 37011 37013 37020 37027 37031 37050 37055 37080 37093 37100 37121 37127 37132 37137 37141 37143 37150 37157 37170 37180 37193 37194 37023 37044 37066 37071 37106 37124 37160 37175 37182 37192 37022 37068 37082 37129 37134 37145 37156 37171 37181 37187 19083 37161 37178 37088 37024 37098 37135 37139 37176 37153 37060 37116 37196 37015 37073 37109 37122 37133 37162 37185 37026 37096 37177 37002 37012 37017 37029 37039 37041 37047 37048 37052 37053 37056 37077 37083 37084 37086 37087 37095 37102 37104 37107 37110 37113 37120 37123 37125 37128 37130 37136 37142 37158 37159 37163 37165 37168 37189 37069 37174 37126

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1855-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.