Z-1871-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- May 29, 2024
- Initiation Date
- April 15, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 18 units
Product Description
Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143
Reason for Recall
Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity
Distribution Pattern
Worldwide - US Nationwide distribution.
Code Information
UDI-DI: (01) 00884838085015 (21) + Serial number Serial Numbers- Applied to FCO72800819: 500046, 500064, 500074, 500102, 500110, 500111, 500145, 500168, 500178, 500211 500216, 500217, 500222, 500225, 500230, 500240, 500242. Serial Number-Applied to FCO72800820 : 500615