Z-1886-2016 Class II Terminated
FDA device recall Z-1886-2016 was initiated by Siemens Healthcare Diagnostics Inc. on Mar 24, 2016 and is designated Class II. Reason for recall: Discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare confirmed a software defect which, in a very specific set of circumstances, results in the Dimension Vista… The recall status is terminated (terminated Mar 13, 2017). Affected quantity: 1300.
Recall Details
- Product Type
- Devices
- Report Date
- Jun 8, 2016
- Initiation Date
- Mar 24, 2016
- Termination Date
- Mar 13, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1300
Product Description
Dimension Vista 1500 Intelligent Lab System running on software versions V.3.6.1 SP1 or V.3.6.2, Device listing # D011374, All serial numbers and lots are affected. The Dimension Vista System is an in vitro diagnostic device intended to duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. Dimension Vista chemical and immunochemical applications use photometric, turbidimetric, chemiluminescence, nephelometric and integrated ion-selective multisensor technology for clinical use
Reason for Recall
Discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare confirmed a software defect which, in a very specific set of circumstances, results in the Dimension Vista System omitting an aliquot probe rinse between sample aspirations when processing tubes in Sample Racks that are front loaded on the Dimension Vista System.
Distribution Pattern
Nationwide Distribution to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, , ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Puerto Rico.
Code Information
All serial numbers are affected.