Z-1895-2024 Class II Ongoing

Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Best

FDA device recall Z-1895-2024 was initiated by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. on May 16, 2024 and is designated Class II. Reason for recall: When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table. The recall status is ongoing. Affected quantity: 1 unit (US).

Recall Details

Product Type
Devices
Report Date
Jun 5, 2024
Initiation Date
May 16, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit (US)

Product Description

Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012, Software Version Number 8.1.100.

Reason for Recall

When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.

Distribution Pattern

Domestic: OK International: India, Spain, U.A.E.

Code Information

Product number: 722012, UDI/DI: 00884838059054; Serial Number: 2138.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1895-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.