Z-1899-2012 Class I Terminated

Recalled by Carefusion 2200 Inc — McGaw Park, IL

FDA device recall Z-1899-2012 was initiated by Carefusion 2200 Inc on May 29, 2012 and is designated Class I. Reason for recall: The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated… The recall status is terminated (terminated Jan 22, 2016).

Recall Details

Product Type
Devices
Report Date
Jul 11, 2012
Initiation Date
May 29, 2012
Termination Date
Jan 22, 2016
Voluntary/Mandated
Voluntary: Firm initiated

Product Description

AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient.

Reason for Recall

The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.

Distribution Pattern

Worldwide distribution: USA (nationwide) including Puerto Rico and internationally to: Canada, Egypt, France, Germany, Great Britain, Italy, Kuwait, Mexico, Portugal, Saudi Arabia, Spain, Taiwan and United Arab Emirates.

Code Information

product codes 10127-4H1, 10331N-4S2, 10351-4H2, 10387-4S2, 10555-4S2, 10706-4S2, 10790-4S2, 10814-4S2, 10849-4S2, 1175-4H2, 1208-4H1, 1553-4H1, 1815-409, 1998-4H1, 2119-4S2G, 2120-4S2G, 4319-4H2, 4326-4H2, 4652-5014677-4H2, 4875-4H2, 5665-4H2, 5682-409, 5732-4H1, 5868-4H2, 6002-4H2, 6040-501, 6074-4H1, 6198-4H2, 6274-H12, 6301-H12, 6313-501, 6459-4H2; 6485-4H2, 6588-4H2, 6942-4H2, 7100-4S2, 7147-4S2, 7156-4S2, 7184-4S2, 7206-4S2, 7213-4S2, 7354-4S2, 7395-4S2, 7407-4S2, 7441-4S2, 7483-4S2, 7526-4S2, 7575-4S2, 7582-4S2, 7588-4S2, 9053-4S2, 9054-4S2, 9055-4S2, 9056-4S2, 9058-4S2, RC41-12027, RC41-18039, RC41-18052, RC41-18060, RC51-12074, RC51-18004, RT4851-00, RT4851-12 and RT4851-18; includes all lot numbers of the affected product codes manufactured prior to February 3, 2012, lot #0000414257 and any lot numbers beginning with the letter Y.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1899-2012
openFDA file date
Oct 27, 2026
Loaded by RegDataLab
Oct 29, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.