Z-1921-2018 Class II Terminated

Recalled by Smith & Nephew, Inc. — Memphis, TN

FDA device recall Z-1921-2018 was initiated by Smith & Nephew, Inc. on May 2, 2018 and is designated Class II. Reason for recall: Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm. The recall status is terminated (terminated Sep 1, 2020). Affected quantity: 13 units.

Recall Details

Product Type
Devices
Report Date
May 30, 2018
Initiation Date
May 2, 2018
Termination Date
Sep 1, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 units

Product Description

smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.

Reason for Recall

Three lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.

Distribution Pattern

US nationwide distribution.

Code Information

Lot Numbers: 17LM22019, 17LM10934, 17LM05900

Source

Dataset
openFDA device recall enforcement reports