Z-1924-2018 Class II Terminated

Recalled by Baxter Healthcare Corporation — Deerfield, IL

Recall Details

Product Type
Devices
Report Date
May 30, 2018
Initiation Date
April 24, 2018
Termination Date
June 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2772 units

Product Description

Prismaflex 7.XX ROW, Product Code 114870 Intended Use: The Prismaflex Control Unit is intended for: " Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. " Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated. " Continuous Renal Replacement Therapy (CRRT) in conjunction with the MARS system to conduct MARS treatments for patients weighing 20 kilograms or more. All treatments administered via the Prismaflex control unit must be prescribed by a physician.

Reason for Recall

Baxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a Voltage Out of Range malfunction alarm, which causes the device to enter a safe state and become inoperable until it is serviced. Baxter will be releasing new firmware that will prevent the malfunction from occurring.

Distribution Pattern

Worldwide distribution: United States (including Puerto Rico), and countries of: Canada, Australia, New Zealand, China, Bangladesh, Japan, Malaysia, Myanmar, Sri Lanka, India, Singapore, Vietnam, Hong Kong, Korea, Taiwan, Indonesia, Bruni, Philippines, Thailand, Argentina, Chile, Costa Rica, Dominican Republic, Haiti, Barbados, Bermuda, Venezuela, Cuba, Mexico, Colombia, Brazil and EMEA.

Code Information

All Lots and Serial Numbers