Z-1926-2016 Class II Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-1926-2016 was initiated by Medtronic Neuromodulation on May 3, 2016 and is designated Class II. Reason for recall: Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a… The recall status is terminated (terminated Aug 4, 2017). Affected quantity: 60,941.

Recall Details

Product Type
Devices
Report Date
Jun 15, 2016
Initiation Date
May 3, 2016
Termination Date
Aug 4, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
60,941

Product Description

Medtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or limbs-including unilateral or bilateral pain.

Reason for Recall

Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two days rather than the typical 30 days. Insufficient coupling between recharger and device during recharge session was found to be key factor.

Distribution Pattern

Worldwide Distribution

Code Information

All lot/serial numbers are affected for both Model numbers 37714 and 97714.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1926-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.