Z-1937-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 12, 2024
- Initiation Date
- April 22, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 25,214,334 syringes
Product Description
MEDLINE LUER LOCK DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101010
Reason for Recall
It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Panama, UAE, USVI.
Code Information
UDI/DI 10884389596319, All lots in the format: 897YYMMXXXX