Z-1948-2014 Class I Terminated

Recalled by Ventlab LLC — Grand Rapids, MI

Recall Details

Product Type
Devices
Report Date
July 9, 2014
Initiation Date
May 16, 2014
Termination Date
November 17, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
251,384

Product Description

Manual Resuscitator Bag Series: AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM1000, CPRM2000, CPRM3000, PRO-1900, PRO-1000, PRO-2000, SC7000, SC8000, SC8020, SC8120, SC9000, SS3200, VN2000, VN3000, VN4000, VN5000, VT1000 , VN2102, VN2002. Emergency manual resuscitator for the Pediatric Population.

Reason for Recall

The pop-off valve in the defective devices remain open and a squeeze of the bag may not generate enough force to force the duck bill valve open and therefore no air will be delivered to the victim.

Distribution Pattern

US Nationwide Distribution in the states of OH, NY, TX, TN, FL, CT, and VA,

Code Information

Manual Resuscitator Bag Series: AF1000, AF2000, AF5000, BT4000, VN2000, VN5000, VT1000; Lot #s105147 -107609. AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, PRO-1900, SC7000, SC8120, SS3200, VN2000, VN3000, VN4000, VN5000; Lot #s 106245 -107291. VN2102; Lot #200349. AF1000, AF2000, AF3000, AF4000, AF5000, BT2000, BT3000, BT4000, BT5000, BVM700, CPRM2000, CPRM3000, PRO-1000, PRO-2000, SC7000, SC8020, VN2000, VN3000, VN4000; Lot #s 107029 -107634. VN2002; Lot # 200492. SC8000, SC9000; Lot #s 101441 -107461. CPRM1000; Lot #s 99523 -107315.