Z-1880-2016 Class II Terminated

Recalled by Ventlab LLC — Grand Rapids, MI

Recall Details

Product Type
Devices
Report Date
June 8, 2016
Initiation Date
March 15, 2016
Termination Date
November 1, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2430

Product Description

STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide gas analyzer Intended for use in various clinical situations to provide respiratory support in the presence of reversible apnea commonly associated with respiratory arrest. The device should only be used by personnel trained in CPR procedures. The ACE/STAT-Check II is intended for use as an adjunct assessment tool for verification of successful ventilation / intubation in patients > 2.2 kilograms (5lbs.).

Reason for Recall

Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during use and will not be an indicator of CO2 level or improper performance of cardiac message and ventilation.

Distribution Pattern

Nationwide distribution to AZ, MS, NY, MO, FL, WI, TX, FL, and CA.

Code Information

Model SC9001B Adult STAT-Check Resuscitation Bag w/ Inflatable Bag Reservoir & Mask Lot numbers 303320 and 303612 Model SC9001C Adult STAT-Check Resuscitation Bag w/ Corrugate Tube Reservoir & Mask Lot number 303400 Model SC9001C-C Adult STAT-Check Resuscitation Bag w/ Corrugate Tube Reservoir, Mask & Straight Connector Lot numbers 303523 and 303791 Model SC8001B Pediatric STAT-Check Resuscitation Bag w/ Inflatable Bag Reservoir & Mask Lot number 303833