Z-1948-2025 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, Inc. — Rochester, NY

Recall Details

Product Type
Devices
Report Date
June 18, 2025
Initiation Date
March 24, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
952 units

Product Description

Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains materials for 48 tests. One kit contains 48 tubes of 1.45 mL Process Buffer and 48 Reaction Tubes. Component: No, not a component

Reason for Recall

It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.

Distribution Pattern

Please see the attached list "Consignee List_CL2025-120." Note: This file contains a specific tab, titled Shipping Summary, which contains the US and ex-US shipping summary. One (1) Distributor was affected. The other four(4) Distributors were sent the notification for information only. QuidelOrtho sent the notification to their consignees because QuidelOrtho drop-shipped this product to their 31 consignees (Included in the Medical Facility total below).

Code Information

Lot Code: Catalog number: M308 UDI-DI: 30014613336167; Lots: 245251 - Exp. 2025-Oct-25 246941 - Exp. 2025-Dec-03 250163 - Exp. 2026-Feb-06 259686 - Exp. 2026-Sep-03 261072 - Exp. 2026-Oct-02 262113 - Exp. 2026-Oct-23 263331 - Exp. 2026-Nov-22 264065 - Exp. 2026-Dec-11 265007 - Exp. 2027-Jan-03 (corrected exp date from 2026-Dec-31) 266928 - Exp. 2026-Dec-31 Additional Lots: 265007-1 - Exp. 2026-Dec-31 268479 and above - Various Expiration dates *Until further notice, all future lots, both supplied by QuidelOrtho or supplied by its distributors, will be affected.