Z-1964-2025 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

Recall Details

Product Type
Devices
Report Date
June 18, 2025
Initiation Date
May 12, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
8027 units

Product Description

Baxter Novum IQ Syringe Pump, product code 40800BAXUS,

Reason for Recall

Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.

Distribution Pattern

US Nationwide distribution.

Code Information

UDI/DI 05413765852428, All Serial Numbers