Z-2017-2013 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-2017-2013 was initiated by Zimmer, Inc. on Jun 7, 2013 and is designated Class II. Reason for recall: Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface… The recall status is terminated (terminated Dec 15, 2014). Affected quantity: 192,355 all devices.

Recall Details

Product Type
Devices
Report Date
Aug 28, 2013
Initiation Date
Jun 7, 2013
Termination Date
Dec 15, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
192,355 all devices

Product Description

90595202017 Articular Surface XLPE CR ART SURF AE12/PUR 17, Rx, Sterile; 90595202020 Articular Surface XLPE CR ART SURF AE12/PUR 20, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.

Reason for Recall

Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the country of Japan.

Code Information

all codes

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2017-2013
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.