Z-2020-2013 Class II Terminated
FDA device recall Z-2020-2013 was initiated by Zimmer, Inc. on Jun 7, 2013 and is designated Class II. Reason for recall: Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface… The recall status is terminated (terminated Dec 15, 2014). Affected quantity: 192,355 all devices.
Recall Details
- Product Type
- Devices
- Report Date
- Aug 28, 2013
- Initiation Date
- Jun 7, 2013
- Termination Date
- Dec 15, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 192,355 all devices
Product Description
00597202009 Articular Surface CR ART SURF 12/PURPLE 09, Rx, Sterile; 00597202010 Articular Surface CR ART SURF 12/PURPLE 10, Rx, Sterile; 00597202012 Articular Surface CR ART SURF 12/PURPLE 12, Rx, Sterile; 00597202014 Articular Surface CR ART SURF 12/PURPLE 14, Rx, Sterile; 00597202017 Articular Surface CR ART SURF 12/PURPLE 17, Rx, Sterile; 00597202020 Articular Surface CR ART SURF 12/PURPLE 20, Rx, Sterile; 00597202023 Articular Surface CR ART SURF 12/PURPLE 23, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
Reason for Recall
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
Distribution Pattern
Worldwide Distribution-USA (nationwide) and the country of Japan.
Code Information
all codes
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2020-2013
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab