Z-2029-2020 Class II Terminated

Recalled by K2M, Inc — Leesburg, VA

Recall Details

Product Type
Devices
Report Date
May 27, 2020
Initiation Date
April 13, 2020
Termination Date
November 15, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90177, UDI # 10888857108905, Size 28x4mm, 18 Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.

Reason for Recall

The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.

Code Information

Lot # FKXV