Z-1410-2021 Class II Terminated

Recalled by K2M, Inc — Leesburg, VA

FDA device recall Z-1410-2021 was initiated by K2M, Inc on Mar 16, 2021 and is designated Class II. Reason for recall: Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of… The recall status is terminated (terminated Aug 12, 2021). Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Apr 21, 2021
Initiation Date
Mar 16, 2021
Termination Date
Aug 12, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the anterior column (cervical spine) Catalog number: 3608-90033

Reason for recall

Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies

Distribution pattern

PA

Code information

Lot Number: MNAJ GTIN: 10888857082571

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1410-2021
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for K2M, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1410-2021. Report a data issue: hello@regdatalab.com.