Z-2031-2025 Class I Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

Recall Details

Product Type
Devices
Report Date
July 16, 2025
Initiation Date
May 30, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9899 units

Product Description

Hillrom Q-Link 13 component, Product Code 3156509, used with the following Mobile Patient Lifts: 1) Uno 102 EE Mobile lift - Product Code 2010004; 2) Viking L Mobile lift - Product Code 2040044; 3) Viking XL Mobile lift - Product Code 2040043; 4) Viking M Mobile lift - Product Code 2040045A; 5) Viking S Mobile Lift - Product Code 2040006; 6) Viking XS Mobile Lift - Product Code 2040007; 7) LikoLight Mobile lift - Product Code 2030001. .

Reason for Recall

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Distribution Pattern

Worldwide

Code Information

All Q-link 13 manufactured between 8/27/2013 to 2/27/2025: 1) Product Code 2010004, UDI/DI 00887761981775; 2) Product Code 2040044, UDI/DI 00887761981867; 3) Product Code 2040043, UDI/DI 00887761981850; 4) Product Code 2040045A, UDI/DI 00887761943193; 5) Product Code 2040006, UDI/DI N/A; 6) Product Code 2040007, UDI/DI N/A; 7) Product Code 2030001, UDI/DI N/A. .