Z-2032-2025 Class I Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 16, 2025
- Initiation Date
- May 30, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 9839 units
Product Description
Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.
Reason for Recall
The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.
Distribution Pattern
Worldwide
Code Information
All LikoScale Adapter Kits manufactured between 8/27/2013 to 2/27/2025