Z-2046-2015 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-2046-2015 was initiated by Biomet, Inc. on Jun 4, 2015 and is designated Class II. Reason for recall: The thread location on one of the thread holes is offset from the axis of the predrill hole. The threads are too deep on one side and too shallow on the other. The locking screw may back out of the… The recall status is terminated (terminated Dec 22, 2015). Affected quantity: 75 units.

Recall Details

Product Type
Devices
Report Date
Jul 15, 2015
Initiation Date
Jun 4, 2015
Termination Date
Dec 22, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
75 units

Product Description

A.L.P.S. Foot Locking Calcaneus Plate,Small-Right. Provides the orthopaedic surgeon a means of bone fixation and helps generally in the management of fractures and reconstructive surgeries.

Reason for Recall

The thread location on one of the thread holes is offset from the axis of the predrill hole. The threads are too deep on one side and too shallow on the other. The locking screw may back out of the plate if it does not achieve proper purchase. This may require a revision surgery to replace or remove the screw and/or plate. The threads, not uniformly loaded, may shear off, causing metal slivers.

Distribution Pattern

Distributed in the states of TX, MD, PA, NJ, NC, MA, AZ, GA, FL, and AR. and the countries of Costa Rica CP and The Netherlands.

Code Information

Part Number: 816209001; Lots: 015840, 124980, 948210 and P0059, P0076, or P0061 will be etched on the plate itself.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2046-2015
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.