Z-2055-2013 Class II Terminated
FDA device recall Z-2055-2013 was initiated by Zimmer, Inc. on Aug 1, 2013 and is designated Class II. Reason for recall: Zimmer is initiating a lot specific recall of the Trabecular Metal" Reverse Glenosphere Distractor due to the potential of the device not functioning properly should it be utilized without proper… The recall status is terminated (terminated Mar 2, 2015). Affected quantity: 771.
Recall Details
- Product Type
- Devices
- Report Date
- Sep 4, 2013
- Initiation Date
- Aug 1, 2013
- Termination Date
- Mar 2, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 771
Product Description
00 4309-049-00 Trabecular Metal Reverse Shoulder Glenosphere Distractor Rx, Non sterile (sterilize before use) The Glenosphere Distractor must be able to attach, connect, or mate with the Glenosphere and remove it without dislodging or loosening the base plate from the glenoid, or damaging the glenoid. Pulling the trigger releases a weight inside the Distractor. This force is transmitted to the bottom of the glenosphere dislodging it from the base plate.
Reason for Recall
Zimmer is initiating a lot specific recall of the Trabecular Metal" Reverse Glenosphere Distractor due to the potential of the device not functioning properly should it be utilized without proper lubrication. As a result, there is a potential for the device to fail to provide impact to the Glenosphere when the trigger is pulled. In addition, there is a potential for the foot of the device to fract
Distribution Pattern
Worldwide Distribution - USA (nationwide) including the countries of Singapore, Brazil, Taiwan, Australia, Canada, Germany(with further distribution to UAE, Austria, Belgium, Switzerland, Czech Republic, Denmark, Spain, Finland, France, Greece, Israel, Italy, Malta, Norway, Portugal, Sweden and South Africa,
Code Information
Item # 00-4309-049-00 and Lot Numbers: 60452884, 60461858, 60531990, 60549559, 60605842, 60612781, 60684768, 60754281, 60815778, 60909547, 60959086, 60978338, 61029048, 61135953, 61213252, 61274877, 61334929, 61431672, 61566464, 61594740, 61690097, 61756058, 61857748, 61858851, 61889174, 61906350, 61910781, 61918666 62039450, 62096379, 62101249, 62141887, 62160965, 62215378, 62254111, 62269856, 62304206, and 62311096.