Z-2063-2016 Class II Terminated

Recalled by Medtest Holdings, Inc. — Canton, MI

FDA device recall Z-2063-2016 was initiated by Medtest Holdings, Inc. on Apr 3, 2015 and is designated Class II. Reason for recall: Deterioration in performance of the Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT resulting in under recovery. The manufacturer has indicated that the HOMOCYSTEINE REAGENT lot listed above shows… The recall status is terminated (terminated Jul 5, 2016). Affected quantity: 13.

Recall details

Product Type
Devices
Report Date
Jul 6, 2016
Initiation Date
Apr 3, 2015
Termination Date
Jul 5, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13

Product description

Liquid stable 2 part Homocysteine Reagent, for in vitro quantitative determination of total homocysteine in human serum and plasma - H7575-57 kit includes one Rl and one R2 bottle and one bottle of calibrator level 1 and one bottle of calibrator level 2.

Reason for recall

Deterioration in performance of the Liquid Stable (LS) 2-Part HOMOCYSTEINE REAGENT resulting in under recovery. The manufacturer has indicated that the HOMOCYSTEINE REAGENT lot listed above shows deterioration in assay performance manifesting itself by under recovery of Homocysteine in Controls and Patient Samples. The cause of this deterioration has been attributed to a loss of reductant activity in Reagent 1. Testing performed indicated that the affected lot is at the limit of acceptable clinical performance. This observation was consistent across the reportable range of the assay. For this reason, do not continue to use affected reagent lot. Test data generated to date will be acceptable where validity control specifiCations have been met.

Distribution pattern

MI, MN, NJ, IL, NY. US Only.

Code information

Model: H7575-57 Lot: 416703 Expiration date: 2016-01

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2063-2016
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtest Holdings, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtest Holdings, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtest Holdings, Inc. →

Source: openFDA record for Z-2063-2016. Report a data issue: hello@regdatalab.com.