Z-2116-2017 Class III Terminated

Recalled by Medtest Holdings, Inc. — Canton, MI

FDA device recall Z-2116-2017 was initiated by Medtest Holdings, Inc. on Feb 15, 2012 and is designated Class III. Reason for recall: A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not… The recall status is terminated (terminated Jun 2, 2017). Affected quantity: 119.168L across All lots.

Recall details

Product Type
Devices
Report Date
May 24, 2017
Initiation Date
Feb 15, 2012
Termination Date
Jun 2, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
119.168L across All lots

Product description

Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.

Reason for recall

A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not consistent with the 510k submission package insert.

Distribution pattern

US Nationwide Distribution

Code information

Catalog numbers: HA461, HA761 , HA961 and lot codes:022204, 034801, 102401, 119503, and 134101.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2116-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medtest Holdings, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medtest Holdings, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medtest Holdings, Inc. →

Source: openFDA record for Z-2116-2017. Report a data issue: hello@regdatalab.com.