Z-2068-2023 Class II Ongoing

Recalled by NeuroLogica Corporation — Danvers, MA

Recall Details

Product Type
Devices
Report Date
July 12, 2023
Initiation Date
June 6, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
48 units

Product Description

OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck Model Number: NL5000

Reason for Recall

1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain contaminated boards, which may impact the ability to power on the device, this could lead to delay for patient treatment

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Australia, Canada, Grand Bahama, India, Italy, Korea, Portugal, Saudi, Scotland, Thailand.

Code Information

GTIN: 10815411020335 10815411020663 Serial Numbers: 10 14 23 25 26 36 38 40 41 42 44 46 48 49 50 51 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 15R