Z-2275-2019 Class II Terminated

Recalled by NeuroLogica Corporation — Danvers, MA

Recall Details

Product Type
Devices
Report Date
August 21, 2019
Initiation Date
July 25, 2019
Termination Date
September 11, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3582 worldwide

Product Description

HS70A Diagnostic Ultrasound System Version 2.01.00, 2.01.01, 2.01.02, 2.01.03, 2.01.04, 2.01.05, 2.01.06, 1.00.06, 1.00.07, 1.00.08, 1.00.09, 1.00.10, 1.00.11, 1.00.12

Reason for Recall

There is a potential for probes overheating when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode.

Distribution Pattern

US Nationwide

Code Information

UDI: 08806167789220 Serial Numbers: S1C0M3HK300004A S1C0M3HK300005P S1C0M3HK300006D S1C0M3HK300007X S1C0M3HK300008N S1C0M3HK300001K S1C0M3HK300002Z S1C0M3HK400005Z S1C0M3HK400006Y S1C0M3HK400008P S1C0M3HK400001L S1C0M3HK400002T S1C0M3HK400003B S15NM3HK200005N S15NM3HK100001X S1C0M3HKC00001H S1C0M3HKC00002J S1C0M3HK900001Z S1C0M3HK900002Y S1C0M3HK900003A S1C0M3HK900004P S1C0M3HK900005D S1C0M3HK900006X S1C0M3HK900007N S1C0M3HK900008E S1C0M3HK900009F S1C0M3HK900010L S1C0M3HK900011T S1C0M3HK800001P S1C0M3HK800002D S1C0M3HK700001N S1C0M3HK700002E