Z-2098-2025 Class II Ongoing

Recalled by GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING — Tirat Carmel, N/A

Recall Details

Product Type
Devices
Report Date
July 16, 2025
Initiation Date
June 20, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
21

Product Description

Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Reason for Recall

Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

Serial Numbers: Millennium VG 1074; "Varicam Millennium VG Discovery VH" 1003 1092 10147 10242 10374 10378 10539 10543 10550 10553 10561 10569 10570 10571 10579 10582 10587 10599 10604 3101-1502