Z-2110-2013 Class II Terminated

Recalled by Becton Dickinson & Co. — Sparks, MD

FDA device recall Z-2110-2013 was initiated by Becton Dickinson & Co. on Jul 8, 2013 and is designated Class II. Reason for recall: PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative. The recall status is terminated (terminated May 1, 2014). Affected quantity: 890 instruments.

Recall Details

Product Type
Devices
Report Date
Sep 4, 2013
Initiation Date
Jul 8, 2013
Termination Date
May 1, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
890 instruments

Product Description

PrepStain AG. For use in the screening and detection of cervical cancer.

Reason for Recall

PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.

Distribution Pattern

Worldwide Distribution-USA (nationwide) and Puerto Rico and the country of Canada.

Code Information

Catalog Numbers - 05CR00021, 05CR00021R, 05CR00022, 05CR00022R, 05CR00023, 05CR00023R