Z-2112-2013 Class II Terminated

Recalled by Becton Dickinson & Co. — Sparks, MD

FDA device recall Z-2112-2013 was initiated by Becton Dickinson & Co. on Jul 8, 2013 and is designated Class II. Reason for recall: PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative. The recall status is terminated (terminated May 1, 2014). Affected quantity: 890 instruments.

Recall Details

Product Type
Devices
Report Date
Sep 4, 2013
Initiation Date
Jul 8, 2013
Termination Date
May 1, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
890 instruments

Product Description

PrepStain Tecan US-II. For use in the screening and detection of cervical cancer.

Reason for Recall

PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.

Distribution Pattern

Worldwide Distribution-USA (nationwide) and Puerto Rico and the country of Canada.

Code Information

Catalog Numbers - 799-14000-00 (490100) and 799-14000-00R (490407)