Z-2131-2014 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-2131-2014 was initiated by Zimmer, Inc. on Jun 10, 2014 and is designated Class II. Reason for recall: A complaint was received in which it was reported that an 8-hole plate from lot 62276042 had an enlarged hole that is unable to lock the screw, which resulted in the screw head pulling through plate… The recall status is terminated (terminated Jan 9, 2015). Affected quantity: 28 units.

Recall Details

Product Type
Devices
Report Date
Aug 6, 2014
Initiation Date
Jun 10, 2014
Termination Date
Jan 9, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28 units

Product Description

Zimmer Periarticular Locking Plate System Distal Radial Volar Plate Lateral Column (1st generation stainless steel only). Orthopedic fixation plate. For temporary internal fixation and stabilization of osteotomies and fractures.

Reason for Recall

A complaint was received in which it was reported that an 8-hole plate from lot 62276042 had an enlarged hole that is unable to lock the screw, which resulted in the screw head pulling through plate hole.

Distribution Pattern

Worldwide Distribution - US including the states of CA and OH., and the countries of BRAZIL, CHINA, KOREA and TAIWAN.

Code Information

Part Number 00-2358-014-04; lot 62276042

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2131-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.