Z-2131-2024 Class II Ongoing

Recalled by B Braun Medical Inc — Breinigsville, PA

Recall Details

Product Type
Devices
Report Date
June 19, 2024
Initiation Date
May 17, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
20,850 units

Product Description

Stimuplex A, 30 DEG, 24GX1", 0.55x25mm, Catalogue Number: 4894251.

Reason for Recall

The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP.

Distribution Pattern

Worldwide distribution - US Nationwide and the country of Canada.

Code Information

Catalogue Number: 4894251; UDI/DI: 04022495101167; Lot/Batch: 20E15, 19L09, 19L27, 20A19, 20H21, 21G14, 21I16, 22H23,