Z-2136-2018 Class II Terminated

Recalled by Roche Diagnostics Corporation — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
June 20, 2018
Initiation Date
April 3, 2018
Termination Date
November 21, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35

Product Description

cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement.

Reason for Recall

The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.

Distribution Pattern

US Distribution to the states of : CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, and VA.

Code Information