Z-2136-2018 Class II Terminated

Recalled by Roche Diagnostics Corporation — Indianapolis, IN

FDA device recall Z-2136-2018 was initiated by Roche Diagnostics Corporation on Apr 3, 2018 and is designated Class II. Reason for recall: The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe… The recall status is terminated (terminated Nov 21, 2019). Affected quantity: 35.

Recall details

Product Type
Devices
Report Date
Jun 20, 2018
Initiation Date
Apr 3, 2018
Termination Date
Nov 21, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35

Product description

cobas e 801 immunoassay analyzer, cobas 8000 Modular Series The device is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement.

Reason for recall

The syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.

Distribution pattern

US Distribution to the states of : CA, CT, IA, IN, KY, ME, MI, NC, NY, OH, SC, TX, and VA.

Code information

Serial numbers: 1744-08 1744-09 1744-10 1758-08 1606-09 1752-05 1752-06 1630-08 1738-01 1744-07 1758-09 1759-04 1758-06 1758-07 1743-09 1743-10 1752-04 1759-05 17C5-06 17C5-07 1615-08 1612-10 17C5-09 1744-05 1744-06 1743-07 1743-08 1744-03 1744-04 1759-06 17C5-05 17C5-08 17C5-10 17C6-01 17C6-02

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2136-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Roche Diagnostics Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2136-2018. Report a data issue: hello@regdatalab.com.