FDA device recall Z-2136-2020 was initiated by Beckman Coulter Inc. on January 29, 2020 and is designated Class II. Reason for recall: Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in th… The recall status is terminated (terminated April 28, 2021). Affected quantity: 2738 vials.
Recall Details
- Product Type
- Devices
- Report Date
- June 3, 2020
- Initiation Date
- January 29, 2020
- Termination Date
- April 28, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2738 vials
Product Description
ISE Electrolyte Reference (2 X 2L) Reagent, REF A28937, UDI: 15099590202491 - Product Usage: are intended for the quantitative determination of carbon dioxide in human serum or plasma.
Reason for Recall
Electrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH and may cause CO2 calibration failures, high CO2 quality control recoveries, and false high patient results which can lead to delay in treatment.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, FL, GA, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the countries of CANADA, JAPAN, LEBANON, SAUDI ARABIA, TAIWAN.
Code Information
Lot Number: M909206