Z-2162-2024 Class II Ongoing

Recalled by Beaver Visitec International, Inc. — Waltham, MA

Recall Details

Product Type
Devices
Report Date
June 26, 2024
Initiation Date
May 10, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
126 units

Product Description

Xstar Safety Slit Knife, 2.75 mm 45¿ Bevel Up, Model Number 378227, a component of CustomEye Kits, Numbers 584562, 5800109

Reason for Recall

The wrong configuration of the blade was in the package. The bevel was on the opposite side, instead of being in front (bevel up) it was in the back (bevel down).

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, UK, and Italy.

Code Information

Xstar Safety Slit Knife: UDI-DI: 00886158001652; Lot Number: 6058854; CustomEye Kit # 584562 , lot # 6068524; CustomEye Kit # 584562 , lot # 6068025; CustomEye Kit # 5800109, lot # 6068285