Z-2170-2012 Class II Terminated

Recalled by Abbott Point Of Care Inc. — Princeton, NJ

Recall Details

Product Type
Devices
Report Date
August 15, 2012
Initiation Date
February 1, 2012
Termination Date
February 8, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,004,200 cartridges US; 54,200 cartridges OUS

Product Description

i-Stat BNP cartridges Abbott Point of Care Inc. The i-Stat BNP test is an in vitro diagnostic test for the quantitative measurement of B-type Natriuretic Peptide (BNP) in whole blood or plasma samples using EDTA as the anticoagulant. BNP measurements can be used as an aid in the diagnosis and assessment of the severity of congestive heart failure.

Reason for Recall

The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.

Distribution Pattern

Worldwide Distribution - USA including AZ and CO and internationally to Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Finland, France, Georgia, Germany, Greece, Honduras, Hong Kong, India, Ireland, Italy, Korea, Kuwait, Mauritius, Mexico, Netherland (Antilles), Nigeria, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, R.O.C., Thailand, UAE, United Kingdom, and Yemen.

Code Information

List number 06F30-01, 06F30-02, 03P93-25, 600-9010-25 All lots