Z-2172-2014 Class II Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-2172-2014 was initiated by Medtronic Neuromodulation on Jul 11, 2014 and is designated Class II. Reason for recall: Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a… The recall status is terminated (terminated Jul 17, 2015). Affected quantity: 1200 (551 US, 649 OUS).

Recall Details

Product Type
Devices
Report Date
Aug 20, 2014
Initiation Date
Jul 11, 2014
Termination Date
Jul 17, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1200 (551 US, 649 OUS)

Product Description

Medtronic Ascenda Intrathecal Catheter models 8780 and 8781, and Ascenda Intrathecal Catheter Pump Segment Revision Kit, model 8784. The Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Ascenda Intrathecal Catheter.

Reason for Recall

Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump. This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria. There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne

Distribution Pattern

Worldwide Distribution - USA (nationwide) including the states of AL,AR,AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, Ml, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WV., and the countries of Austria, Belgium, Cyprus, Denmark, France, Germany, Iraq, Italy, Lebanon, Netherlands, Panama, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, United Kingdom and Uruguay.

Code Information

Serial Numbers are in the following format: "N" + 9 digits (e.g. N12345001. Product beginning with the following letter/number combinations are affected: N464184, N464200, N464206, N464254, N464257, N464258, N464262, N464263, N464264, N464271, N464693, N464703, N464708, N464725, N464727, N464730, N464734, N464736, N464738, N464742, N464746, N464750, N464752, N464841, N464924, N464930, N464995, N464995, N465020, N465038, N465063, N465065, N465073, N465082, N465088, N465091, N465094, N465111, N465125, N465129, N465131, N465136, N465138, N465144, N465145, N465146, N465148, N465150, N466065, N466067, N466067, N466068, N466069, N466071, N466072, N466075, N466134, N466167, N466177, N466196, N466201, N466240, N466244, N466259, N466289, N466290, N466291, N466306, N466368, N466390, N466392, N466394, N466398, N466399, N466631, N466639, N466645, N466678, N466678, N466754, N466758, N466761, N466886, N466897, N466901, N466902, N466911, N466914, N467103, N467109, N467113, N467114, N467116, N467126, N467138, N467171, N467180, N467183, N467192, N467193, N467194, N467195, N467197, N467198, N467203, N467204, N467206, N467209, N467351, N467351, N467352, N467355, N467356, N467357, N467395, N467399, N467401, N467402, N467407, N467444, N467451, N467453, N467454, N467563, N467568, N467570, N467601, N467606, N467608, N467609, N467611, N467612, N467614, N467618, N467621, N467626, N468288, N468343, N468401, N468404, N468406, N468406, N468448, N468494, N468500, N468507, N468510, N468511, N468514, N468516, N468519, N468522, N468574, N468580, N468621, N468622, N468624, N468626, N468641, N469182, N469262, N469332, N469360, N469360, N469388, N469420, N469422, N469428, N469431, N469433, N469434, N469436, N469438, N469440, N469446, N469447, N469449, N469452, N469460, N469465, N469466, N469469, N469474, N469478, N469507, N469511, N469530, N469533, N469549, N469579, N469583, N469586, N469588, N470163, N470197, N470234, N470303, N470307, N470401, N470408, N470414, N470420, N470424, N470453, N470459, N470919, N470923, N470929, N470931, N470931, N470934, N470938, N470940, N470941, N470951, N470953, N470954, N471442, N471443, N471446, N471447, N471448, N471449, N471487, N471488, N471490, N471490, N471491, N471764, N471791, N471840, N471853, N471856, N471860, N471863, N471864, N471865, N471866, N471870, N471871, N471875, N471880, N472310, N472326, N472333, N472336, N472342, N472345, N472347, N472857, N472861, N472866, N472868, N472873, N472875, N472876, N472878, N472882, N472885, N472886, N472890, N472894, N472900, N472905, N473140, N473144, N473148, N473154, N473156, N473176, N473178, N473180, N473202, N473207, N473211, N473221, N473222, N473223, N473228, N473230, N473311, N473311, N473323, N473377, N473401, N473426, N473438, N473492, N473493, N473989, N473996, N473999, N474001, N474005, N474006, N474007, N474010, N474018, N474019, N474023, N474027, N474037, N474038, N474047, N474047, N474421, N474423, N474427, N474431, N474433.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2172-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.