Z-2181-2018 Class II Terminated

Recalled by Siemens Healthcare Diagnostics, Inc. — Newark, DE

Recall Details

Product Type
Devices
Report Date
June 20, 2018
Initiation Date
April 19, 2018
Termination Date
April 24, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7543

Product Description

Dimension Gentamicin Flex reagent cartridge, DM GENT, gentamicin enzyme immunoassay, Lot numbers affected are EB8115, BB8159, and FB8275.

Reason for Recall

The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators.

Distribution Pattern

Nationwide distribution. Foreign distribution to Argentina, Australia, Brazil, Bahamas, Canada, Germany, India, Japan, Myanmar, New Zealand, Saudi Arabia, Taiwan, and South Africa.

Code Information

Lots EB8115 BB8159 FB8275