Z-2182-2024 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
July 3, 2024
Initiation Date
May 31, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product Description

AXIOM MULTIX MP - Intended Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices Material Number: 8395381

Reason for Recall

Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI: N/A Serial Numbers: 1022 1074 1118 1119 1120 1121 1123 1126 1130 1131 1134 1138