Z-2184-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- July 3, 2024
- Initiation Date
- May 31, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2 units
Product Description
AXIOM Vertix MD - The Vertix MD Trauma has been specially designed for examining emergency and accident patients as well as for use in outpatient departments. Material Number: MD39543
Reason for Recall
Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire
Distribution Pattern
US Nationwide distribution.
Code Information
UDI-DI: N/A Serial Numbers: 1040 1020