Z-2185-2021 Class II Ongoing

Recalled by Abbott Laboratories — Irving, TX

FDA device recall Z-2185-2021 was initiated by Abbott Laboratories on Jun 10, 2021 and is designated Class II. Reason for recall: There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with… The recall status is ongoing. Affected quantity: 1,603 units.

Recall details

Product Type
Devices
Report Date
Aug 11, 2021
Initiation Date
Jun 10, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,603 units

Product description

REF 8L92-22, Abbott Activated Alanine aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161859

Reason for recall

There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of VA, UT, TX, TN, SD,PA, NY, ME, ID, CA and the countries of AFGHANISTAN, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVI, CANADA, CAYMAN ISLANDS, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FAROE ISLAND, FINLAND, FRANCE, GERMANY, GREENLAND, Hong Kong, INDIA, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, MONTENEGRO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, PORTUGAL, PR CHINA, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UNITED KINGDOM, VIETNAM.

Code information

Lot Numbers / Expiration Date 02954UN20 / 29AUG2021 37921UN20 / 01DEC2021 81823UN21 / 16FEB2022

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2185-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Laboratories? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Abbott Laboratories: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Abbott Laboratories →

Source: openFDA record for Z-2185-2021. Report a data issue: hello@regdatalab.com.