Z-2187-2024 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-2187-2024 was initiated by Siemens Medical Solutions USA, Inc on May 31, 2024 and is designated Class II. Reason for recall: Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire The recall status is ongoing. Affected quantity: 3 units.

Recall Details

Product Type
Devices
Report Date
Jul 3, 2024
Initiation Date
May 31, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product Description

MULTIX TOP - Intended Use: The Multix TOP I PRO Radiographic X ray Table is a patient table used in conjunction with the Siemens Multix PRO and TOP Systems radiographic devices Material Number: 475525

Reason for Recall

Short circuit in the power supply unit (PSU) of the Touch Display may lead to overheating and fire

Distribution Pattern

US Nationwide distribution.

Code Information

UDI-DI: N/A Serial Numbers: 1523 1585 1897