Z-2211-2024 Class II Ongoing

Recalled by Medtronic Inc. — Mounds View, MN

FDA device recall Z-2211-2024 was initiated by Medtronic Inc. on Apr 16, 2024 and is designated Class II. Reason for recall: Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing. The recall status is ongoing. Affected quantity: 9 units.

Recall Details

Product Type
Devices
Report Date
Jul 3, 2024
Initiation Date
Apr 16, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9 units

Product Description

CROME VR MRI SureScan, Model Number DVPC3D1, Implantable Cardioverter Defibrillator

Reason for Recall

Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of CA, CT, FL, IN, MA, MI, NC, NH, NJ, NM, NY, OH, TX, VA and the countries of Australia, Austria, Bulgaria, Canada, China, Croatia, Czech Republic, France, Georgia, Germany, Greece, Guadeloupe, Iceland, Ireland, Italy, Jordan, Kosovo, Latvia, Moldova, Republic Of, Netherlands, Northern Ireland, Norway, Oman, Panama, Poland, Portugal, Qatar, Reunion, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.

Code Information

GTIN: 00763000178543, Serial Numbers: RSG601442S, RSG601443S, RSG601445S, RSG601446S, RSG601449S, RSG601461S, RSG601467S, RSG601478S; GTIN: 00763000711382, Serial Number: RSG601428S.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2211-2024
openFDA file date
Oct 6, 2026
Loaded by RegDataLab
Oct 12, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it. FDA may have changed the recall since.