Z-2212-2016 Class I Terminated

Recalled by Carefusion 2200 Inc — Vernon Hills, IL

FDA device recall Z-2212-2016 was initiated by Carefusion 2200 Inc on Jul 1, 2016 and is designated Class I. Reason for recall: BD, formerly CareFusion, has identified a potential risk associated with a weld failure which could result in the wire protruding thru the tip of the instrument when articulated. If this failure were… The recall status is terminated (terminated Mar 29, 2017). Affected quantity: 10.

Recall Details

Product Type
Devices
Report Date
Aug 3, 2016
Initiation Date
Jul 1, 2016
Termination Date
Mar 29, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10

Product Description

DIAMOND-FLEX CIRCULAR RETRACTOR, ANGLED, 40MM 5MM, Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16. Designed to retract or elevate organs and tissue to provide better visualization access.

Reason for Recall

BD, formerly CareFusion, has identified a potential risk associated with a weld failure which could result in the wire protruding thru the tip of the instrument when articulated. If this failure were to occur while in use in a procedure it has the potential to damages tissue or organs

Distribution Pattern

Worldwide Distribution - US, including the states of WA and MD; and, the country of Japan.

Code Information

Product Code 89-6114, Lot Code(s): 878971; 879366; 879404, and Date Code: C16

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2212-2016
openFDA file date
Oct 26, 2026