Z-2214-2024 Class II Ongoing

Recalled by Siemens Healthcare Diagnostics Inc — Norwood, MA

FDA device recall Z-2214-2024 was initiated by Siemens Healthcare Diagnostics Inc on May 17, 2024 and is designated Class II. Reason for recall: Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With… The recall status is ongoing. Affected quantity: 9 units US.

Recall Details

Product Type
Devices
Report Date
Jul 3, 2024
Initiation Date
May 17, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9 units US

Product Description

Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e Blood Gas System (Japan) 11416754; RAPIDPoint 500e Blood Gas System (ROW) 11416755 ;

Reason for Recall

Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of FL, GA, LA, MI, NC and the countries of Chile, Costa Rica, France, Germany, Japan, Lebanon, Mexico, Netherlands, New Zealand, Poland, Spain, Turkey, United Kingdom.

Code Information

UDI-DI: (1)RAPIDPoint 500e Blood Gas System (USA) 00630414286143; (2)RAPIDPoint 500e Blood Gas System (China) 00630414286150; (3) RAPIDPoint 500e Blood Gas System (Japan) 00630414286167; (4) RAPIDPoint 500e Blood Gas System (ROW) 00630414286174 Software Version 5.3.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2214-2024
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.