Z-2234-2018 Class II Terminated
FDA device recall Z-2234-2018 was initiated by Zimmer Biomet, Inc. on May 16, 2018 and is designated Class II. Reason for recall: The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires… The recall status is terminated (terminated Apr 5, 2024). Affected quantity: 449.
Recall Details
- Product Type
- Devices
- Report Date
- Jun 27, 2018
- Initiation Date
- May 16, 2018
- Termination Date
- Apr 5, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 449
Product Description
GUIDE WIRE w/ BALL TIP 2.0MMX98CM; 469040 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.
Reason for Recall
The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.
Distribution Pattern
Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.
Code Information
228829 311230 311260 255105 152590 757210 982070 282057 283729 057650 292980 293094 299420 668040 579470 354970 519980 670820 695640 963100
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2234-2018
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab
- Oct 25, 2026