Z-2235-2020 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
June 24, 2020
Initiation Date
April 28, 2020
Termination Date
September 23, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,883 units total

Product Description

Giraffe OmniBed with installed Servo Oxygen module. Infant incubator and warmer.

Reason for Recall

Potential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.

Distribution Pattern

Worldwide Distribution. US nationwide including US Virgin Islands, Australia, Austria, Bahrain, Belgium, Bermuda, Bulgaria, Burundi, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Guernsey, Hong Kong, India, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Republic of Korea, Kuwait, Latam, Latvia, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, OMAN, State of Palestine, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Yemen

Code Information

Giraffe Incubator: serial numbers HDHA through HDHQ; Servo Oxygen Board spare part kit: 6600-0233-850; Servo Oxygen upgrade kits: 6600-0678-800, 6600-0678-801, 6600-0678-802, 6600-0678-803, 66000849-800, 6600-0849-801, 6600-0849-802, 6600-0849-803, 6600-0866-800, 6600-0866-801, 6600-0867800, 6600-0867-801, M1142293 or M1142295.