Z-2237-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
June 27, 2018
Initiation Date
May 16, 2018
Termination Date
April 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
832

Product Description

G-PIN 2.8BL 3.2PL 98CM 2PK; 467260 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.

Reason for Recall

The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.

Distribution Pattern

Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.

Code Information

286530 300210 413620 448170 548100 570250 626810 638980 700950 758900 828020 828040 932250 244880 244890 125970 150380 298880 754180 571170 198490 800020 399250