Z-2243-2020 Class II Terminated

Recalled by Med Tec Inc — Orange City, IA

Recall Details

Product Type
Devices
Report Date
June 10, 2020
Initiation Date
May 12, 2020
Termination Date
March 25, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
171 units

Product Description

Type S Extension For use with Varian, CIVCO, Prescription Only - Product Usage: Device is intended for patient positioning.

Reason for Recall

Potential for disengagement of the Varian Style Type-S Extension during patient repositioning. The risk associated with the disengagement of the extension could be a patient fall if the patient is not supported by clinical staff.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of WI, TX, AZ, FL, NE, TN, MA, FL, GA, IN, OR, PA, IL, LA, NH, MS, MI, OH, MN, VA, IA, CA, NC, SC, AK, UT, WA, MO, MT, KS, and WV. The countries of Africa, Colombia, Saudi Arabia, Costa Rica, South Korea, Hungary, Spain, and Italy.

Code Information

Part Number: MTCFHNV UDI: 00841439104156 Lot Numbers: M027990; M505880; M847500; M166160; M505890; M858580; M198170; M505900; M858590; M241650; M505910; M869030; M249180; M511240; M885500; M257630; M530920; M918000; M278760; M541350; M919980; M282330; M553300; M919990; M283380; M553720; M928400; M293670; M553740; M928410; M300750; M596220; M934260; M303010; M596230; M936110; M314410; M616690; M942400; M321880; M616700; M942510; M323560; M649630; M944110; M326580; M649670; M951030; M333420; M649770; M964730; M350700; M715520; M979830; M356130; M715540; M987460; M362290; M743010; M995860; M365740; M767820; M369700; M774760; M371990; M782390; M418080; M796350; M436620; M796360; M439700; M802120; M442410; M805650; M446460; M809050; M448260; M809060; M457870; M820300; M457880; M829290; M462640; M834400; M493710; M844990; M495030; M845000